Accuracy of Medical Oxygen Flowmeters: A Multicentric Field Study
août 2014, DUPREZ, Frédéric; Barile, Maria; Bonus, Thierry; CUVELIER, Etienne; Ollieuz, Sandra; Mashayekhi, Shahram; Legrand, Alexandre,
- HE Condorcet
,
- Article scientifique
Article scientifique
The accuracy of 476 oxygen flowmeters was investigated using a thermal mass flowmeter in eight hospitals in France and Belgium. Different oxygen flow rates (2 to 15 l/min) were evaluated at the patient’s bed. When the sample was considered as a whole, the accuracy of delivered flow was acceptable but precision was poor. The variability of the delivered flow between devices was greater when a low ...