dc.description.abstract | The accuracy of 476 oxygen flowmeters was investigated using a thermal mass flowmeter in eight hospitals in France and Belgium. Different oxygen flow rates (2 to 15 l/min) were evaluated at the patient’s bed. When the sample was considered as a whole, the accuracy of delivered flow was acceptable but precision was poor. The variability of the delivered flow between devices was greater when a low flow rate was required. Compensated-pressure oxygen flowmeters for these low rates were more accurate than their non-compensated counterparts. This study emphasizes the need to individually adapt the oxygen flow rate each time a patient has to move from one flowmeter to another. | en_US |
dc.language.iso | EN | en_US |
dc.publisher | Scientific Research Publishing | en_US |
dc.relation.ispartof | Health | en_US |
dc.rights.uri | https://creativecommons.org/licenses/by/4.0/ | en_US |
dc.subject | Medical device accuracy | en_US |
dc.subject | Oxygen therapy | en_US |
dc.subject | Flowmeter | en_US |
dc.title | Accuracy of Medical Oxygen Flowmeters: A Multicentric Field Study | en_US |
dc.type | Article scientifique | en_US |
synhera.classification | Sciences de la santé humaine>>Médecine de laboratoire & technologie médicale | en_US |
synhera.classification | Sciences de la santé humaine>>Systèmes cardiovasculaire & respiratoire | en_US |
synhera.institution | HE Condorcet | en_US |
synhera.otherinstitution | Department of Physiology and Pharmacology, University of Mons, Belgium | en_US |
synhera.otherinstitution | Centre Hospitalier Epicura, Hornu | en_US |
synhera.cost.total | / | en_US |
synhera.cost.apc | / | en_US |
synhera.cost.comp | / | en_US |
synhera.cost.acccomp | / | en_US |
dc.description.version | Oui | en_US |
dc.rights.holder | Cuvelier et al. | en_US |